Approval of 6i24r1 et al - E.Coli CIP NSF 6
Comment Submitted by
Jonathan Brania
2025-08-25 16:35:15

If the revisions are meant to ensure that the inoculum density in the machine is what was verified on the bench, then I support the intent, but I would recommend revisions.

  • Dispense from instead of dispense through.  "Dispense through" is less clear if the intent is to collect samples.
  • Add "all" to the 1st sentence and delete "ensure the entire food contact flow path is exposed to sufficient challenge" from the 3rd sentence.  The act of dispensing samples does not ensure that all food contact surfaces were exposed to the inoculum. 
  • Add a sentence at the end to compare Ni to the intended challenge density.  
Submitter Proposed Solution

6.1.2.2 The equipment shall be operated so that all food contact surfaces are exposed to the E. coli and product mix suspension. If a remote product supply system is being tested, the remote line set shall be filled with E. coli and product mix suspension so that all food contact surfaces are exposed (i.e., no air in remote line set). E. coli and product mix suspension shall be dispensed through from the equipment, collecting three 200-mL positive control samples from the dispense point to ensure the entire food contact flow path is exposed to sufficient challenge. The average of the positive control samples shall serve as the initial inoculum density (Ni).  Ni shall be within the range of required density of E. coli culture suspension.